International Journal of Clinical Pediatric Dentistry

Register      Login

VOLUME 17 , ISSUE S1 ( April, 2024 ) > List of Articles

ORIGINAL RESEARCH

Evaluation of Oxygen Saturation Levels Using a Custom-modified Finger Pulse Oximeter for Assessment of Pulp Vitality in Various Clinical Situations in Pediatric Dental Practice: An In Vivo Study

Abhijeet Patil, Nishita Garg, Lumbini Pathivada, Rishika Choudhary, Harsimran Kaur, Ramakrishna Yeluri

Keywords : Oxygen saturation, Pulse oximeter, Pulp vitality

Citation Information :

DOI: 10.5005/jp-journals-10005-2744

License: CC BY-NC 4.0

Published Online: 30-07-2024

Copyright Statement:  Copyright © 2024; The Author(s).


Abstract

Background: A major drawback of traditional pulp testing procedures is that they depend on neural responses rather than vascular circulation. Hence, this study aimed to develop a custom-modified dental sensor using a finger pulse oximeter (PO) that is applicable to any type of tooth so as to test its ability and accuracy in evaluating the oxygen saturation (SaO2) values of teeth at different developmental stages as a measure of pulp vitality. Materials and methods: A customized finger PO was employed to determine the systemic and pulp SaO2 levels in 300 children. A total of 600 teeth (primary and permanent) were divided into group I (100 primary molars), group II (200 permanent molars), group III (200 permanent incisors), and 100 endodontically treated (primary and permanent) teeth were included in group IV. The mean SaO2 values thus obtained in various groups were compared. Results: Intragroup comparison of mean SaO2% of the patient's finger (systemic), test teeth, and control teeth showed significant differences (p ≤ 0.001). Intergroup comparisons also revealed significant differences in most of the groups. Conclusions: The modified PO probe can be applied to any type of tooth. It was found to be accurate and sensitive enough to detect changes in SaO2 levels in various clinical situations irrespective of developmental stage. It proved to be a direct method of evaluation of pulp vitality by objective means.


HTML PDF Share
  1. Noblett WC, Wilcox LR, Scamman F, et al. Detection of pulpal circulation in vitro by pulse oximetry. J Endod 1996;22(1):1–5. DOI: 10.1016/S0099-2399(96)80226-3
  2. Walton RE, Torabinejad M. Principles and Practice of Endodontics. Philadelphia: WB Saunders: 1989. pp. 61–62.
  3. Rowe AH, Pitt Ford TR. The assessment of pulpal vitality. Int Endod J 1990;23(2):77–83. DOI: 10.1111/j.1365-2591.1990.tb00843.x
  4. Andreasen FM, Zhijie Y, Thomsen BL, et al. Occurrence of pulp canal obliteration after luxation injuries in the permanent dentition. Endod Dent Traumatol 1987;3(3):103–115. DOI: 10.1111/j.1600-9657.1987.tb00611.x
  5. Kennedy DCM, Kiely MC, Keating PJ. Efficacy of electrical pulp testing. J Irish Dental Assoc 1987;33(2-4):41–46.
  6. Rapp R, Avery JK, Strachan DS. The distribution of nerves in human primary teeth. Anat Rec 1967;159(1):89–103. DOI: 10.1002/ar.1091590113
  7. Radhakrishnan S, Munshi AK, Hegde AM. Pulse oximetry: a diagnostic instrument in pulpal vitality testing. J Clin Pediatr Dent 2002;26(2):141–145. DOI: 10.17796/jcpd.26.2.2j25008jg6u86236
  8. Gazelius B, Olgart L, Edwall B, et al. Non-invasive recording of blood flow in human dental pulp. Endod Dent Traumatol 1986;2(5):219–221. DOI: 10.1111/j.1600-9657.1986.tb00148.x
  9. Schnettler JM, Wallace JA. Pulse oximetry as a diagnostic tool of pulpal vitality. J Endod 1991;17(10):488–490. DOI: 10.1016/S0099-2399(06)81795-4
  10. Wallace JA, Schnettler JM. Pulse oximetry as a definitive diagnostic tool. J Endod 1993;19:186.
  11. Caldeira CL, Barletta FB, Ilha MC, et al. Pulse oximetry: a useful test for evaluating pulp vitality in traumatized teeth. Dent Traumatol 2016;32(5):385–389. DOI: 10.1111/edt.12279
  12. Sharma DS, Mishra S, Banda NR, et al. In vivo evaluation of customized pulse oximeter and sensitivity pulp tests for assessment of pulp vitality. J Clin Pediatr Dent 2019;43(1):11–15. DOI: 10.17796/1053-4625-43.1.3
  13. Hiremath SS. Biostatistics. Hiremath SS Textbook of Preventive and Community Dentistry. India: Elsevier: 2011. pp. 511–520.
  14. Johnsen DC, Harshbarger J, Rymer HD. Quantitative assessment of neural development in human premolars. Anat Rec 1983;205(4):421–429. DOI: 10.1002/ar.1092050407
  15. Pupim D, Iwaki Filho L, Takeshita WM, et al. Evaluation of accuracy of portable fingertip pulse oximeter, as compared to that of a hospital oximeter with digital sensor. Indian J Dent Res 2013;24(5):542–546. DOI: 10.4103/0970-9290.123362
  16. Goho C. Pulse oximetry evaluation of vitality in primary and immature permanent teeth. Pediatr Dent 1999;21(2):125–127.
  17. Gopi Krishna V, Kandaswamy D, Gupta T. Assessment of the efficacy of an indigeniously developed pulse oximeter dental sensor holder for pulp vitality testing. Indian J Dent Res 2006;17(3):111–113. DOI: 10.4103/0970-9290.29880
  18. Gopikrishna V, Tinagupta K, Kandaswamy D. Comparison of electrical, thermal, and pulse oximetry methods for assessing pulp vitality in recently traumatized teeth. J Endod 2007;33(5):531–535. DOI: 10.1016/j.joen.2007.01.014
  19. Calil E, Caldeira CL, Gavini G, et al. Determination of pulp vitality in vivo with pulse oximetry. Int Endod J 2008;41(9):741–746. DOI: 10.1111/j.1365-2591.2008.01421.x
  20. Fein ME, Gluskin AH, Goon WW, et al. Evaluation of optical methods of detecting dental pulp vitality. J Biomed Opt 1997;2(1):58–73. DOI: 10.1117/12.261679
  21. Pozzobon MH, de Sousa Vieira R, Alves AM, et al. Assessment of pulp blood flow in primary and permanent teeth using pulse oximetry. Dent Traumatol 2011;27(3):184–188. DOI: 10.1111/j.1600-9657.2011.00976.x
  22. Bargrizan M, Ashari MA, Ahmadi M, et al. The use of pulse oximetry in evaluation of pulp vitality in immature permanent teeth. Dent Traumatol 2016;32(1):43–47. DOI: 10.1111/edt.12215
  23. Setzer FC, Kataoka SH, Natrielli F, et al. Clinical diagnosis of pulp inflammation based on pulp oxygenation rates measured by pulse oximetry. J Endod 2012;38(7):880–883. DOI: 10.1016/j.joen.2012.03.027
  24. Wong JK. Pulse oximetry: technology and critical uses. Ont Dent 2016;2:22–27.
  25. Karayilmaz H, Kirzioğlu Z. Evaluation of pulpal blood flow changes in primary molars with physiological root resorption by laser Doppler flowmetry and pulse oximetry. J Clin Pediatr Dent 2011;36(2):139–144. DOI: 10.17796/jcpd.36.2.3l9lgu5292r08742
  26. Batchelder PB, Raley DM. Maximizing the laboratory setting for testing devices and understanding statistical output in pulse oximetry. Anesth Analg 2007;105(6 Suppl):S85–S94. DOI: 10.1213/01.ane.0000268495.35207.ab
PDF Share
PDF Share

© Jaypee Brothers Medical Publishers (P) LTD.